MDO-2000A series User Manual
234
Auto Tolerance
8. To set the boundary tolerance as a
percentage offset from the source
waveform, press Auto Tolerance
and use the Variable knob.
VARIABLE
Offset
0.4% ~ 40% (.4% steps)
Maximum and
Minimum
Position
9. To manually set the template
tolerance, press Minimum Position
or Maximum Position and use the
Variable knob to set the absolute
minimum or maximum position.
or
Range
Voltage division range
Save Boundary
Template
10. Press Save Operation to save the
tolerance boundaries.
11. The Maximum Position tolerance will be saved
to reference waveform R1, and the Minimum
Position tolerance to R2.
12. Press Go Back to return to the
previous menu.
Start Go-NoGo
Press Enable to start the Go-NoGo test. Then the
Enable button will change to Disable. Pressing
Disable will stop the Go-NoGo test and toggle the
button back to Enable.
If the Violating setting was set to Stop, press Enable
to restart the test after it has stopped.