Specifications
Meter Measurements for Fluke 190-xx2
8
119
Temperature (TEMP)
With Optional Temperature Probe
Ranges (°C or °F) ..................................-40.0 to +100.0 °
-100.0 to +250.0 °
-100.0 to +500.0 °
-100 to +1000 °
-100 to + 2500 °
Probe Sensitivity ............................ 1 mV/°C and 1 mV/°F
DC Voltage (VDC)
Ranges .....500.0 mV, 5.000 V, 50.00 V, 500.0 V, 1100 V
Full Scale Reading........................................5000 counts
Accuracy ............................................ ±(0.5 % +5 counts)
Normal Mode AC Rejection at 50 or 60 Hz ±1 % .... >60 dB
AC Voltage (VAC)
Ranges .....500.0 mV, 5.000 V, 50.00 V, 500.0 V, 1100 V
Full Scale Reading........................................5000 counts
Accuracy
15 Hz to 60 Hz ................................. ±(1 % +10 counts)
60 Hz to 1 kHz .............................. ±(2.5 % +15 counts)
For higher frequencies the frequency roll off of the
Meter input starts affecting accuracy.
Normal Mode DC Rejection .................................. >50 dB
AC+DC Voltage (True RMS)
Ranges......500.0 mV, 5.000 V, 50.00 V, 500.0 V, 1100 V
Full Scale Reading ........................................5000 counts
Accuracy
DC to 60 Hz...................................... ±(1 % +10 counts)
60 Hz to 1 kHz............................... ±(2.5 % +15 counts)
For higher frequencies the frequency roll off of the Meter
input starts affecting accuracy.
All accuracies are valid if the waveform amplitude is
larger than 5 % of full scale.
Amperes (AMP)
With Optional Current Probe or Current Shunt
Ranges............................. same as VDC, VAC, VAC+DC
Probe Sensitivity..................100 µV/A, 1 mV/A, 10 mV/A,
100 mV/A, 1 V/A, 10 V/A, and 100 V/A
Accuracy .......................... same as VDC, VAC, VAC+DC
(add current probe or current shunt accuracy)