4.Advanced User Guidebook
83
Rule
Source
Horizontal
Vertical
Create
Select the source as CH1, CH2, CH3 or CH4
Change the Horizontal tolerance value by
turning the M knob
Change the Vertical tolerance value by
turning the M knob
Use the rule set as testing rule
SaveRule
Number
Save
Load
Select any one from Rule1 - Rule8 as your
rule name
Select Save to save the rule
Load some rule as the testing rule
The description of Pass/fail Menu is shown as the follows:
Pass/Fail test:
Detect whether the input signal is within the limits of the rule, if it exceeds limits
of the rule, it is "Fail"; otherwise it is "Pass". Also it can output fail or pass
signal by built-in and configurable output port. To run the test, read the
following steps:
1. Push the Utility button, select Function in the bottom menu, select
Pass/fail in the left menu.
2. Enable switch on: Select Operate in the bottom menu, select Enable in
the right menu as ON.
3. Create rule: Select Rule in the bottom menu. Select Source in the right
menu, select the source in the left menu. Set Horizontal tolerance and
Vertical tolerance in the right menu. Select Create in the right menu to
create the rule.
4. Set output type:
1) Select Output in the bottom menu to enter output settings.
2) Set Type in the right menu to Pass or Fail.
3) When the Stop in the right menu is set to ON, and the measured signal
meets the set rules, the acquisition will stop.
4) When the Beep in the right menu is set to ON, and the measured
signal meets the set rules, the buzzer will beep.
5. Begin to test: Select Operate in the bottom menu, select Operate in the
right menu as Start, the test will begin.
6. Save rule: Select SaveRule in the bottom menu. Select the save location
in the left menu, and then select Save in the right menu to save the rules,
which could be called up at once when need. Select Load to call up the rule
saved.
Note:
1. When Pass/Fail is ON, if XY or FFT is ready to run, then Pass/Fail will be
closed; under the mode of XY or FFT, Pass/Fail is unable.
2. Under the mode of Factory, Auto Scale and Auto Set, Pass/Fail will be
closed.
3. When no save setting left in the rule save, tip will be given to show "NO