Tripp Lite IS600HGDV Isolator Series Dual-Voltage 115/230V 600W 60601-1 Medical-Grade Isolation Transformer, C14 Inlet, 6 C13 Outlets

Owner's Manual - Page 4

For IS600HGDV. Series: AG-8799
Also, The document are for others Tripp Lite models: IS300HGDV, IS1000HGDV, IS1800HGDV

PDF File Manual, 32 pages, Read Online | Download pdf file

IS600HGDV photo
Loading ...
Loading ...
Loading ...
4
3. Installation
3.2 Plug the isolation transformer into an electrical outlet.
Note: The isolation transformer does not include an input power cord (power cords available at www.tripplite.com).
Connect a user-supplied input power cord to the IEC-320-C14 input receptacle for models IS300HGDV, IS600HGDV and IS1000HGDV. Connect
to the IEC-320-C20 input receptacle for model IS1800HGDV. The power cord should have an IEC-320-C13 connector for models IS300HGDV,
IS600HGDV, IS1000HGDV and an IEC-320-C19 connector for IS1800HGDV on one end and a plug appropriate for your local site’s utility outlet
on the other end.
3.3 Plug equipment into the isolation transformer’s AC outlets.
Note: The isolation transformer only meets CAN/ES/IEC/EN 60601-1 when plugged into a properly installed hospital-grade outlet.
3.4 Turn the isolation transformer ON to provide clean AC power to its receptacles.
“|” is the ON position; “O” is the OFF position.
3.1.3 For IS1800HGDV:
Input voltage selection is made via the selector switch labeled “Input Voltage”
A
(Figure 3). Separate fuse holders are available for the
IS1800HGDV. You will have to insert the correct fuse value depending on input voltage. For correct fuse value, see the table in section
3.1.5. Output voltage selection is made with the output voltage selector switch
B
.
Note: If using as a step-up or step-down transformer, the output voltage must match the connected equipment load requirements.
Figure 3
B
Output Voltage
Selector Switch
Use a small, at-blade screwdriver to open the access cover for the
fuses and to set the input and output voltages.
A
3.1.4 For all models:
Output voltage selection is made via the selector switch labeled “Output Voltage”.
3.1.5 Fuse Values and Unit Specications
Model Agency Series Input / Output Input Output 115V Fuse Value 230V Fuse Value
IS300HGDV AG-8798 115/230VAC 50/60Hz 3.15/1.6A 2.6/1.3A 3.15 1.6
IS600HGDV AG-8799 115/230VAC 50/60Hz 6.0/3.15A 5.2/2.6A 6.3 3.15
IS1000HGDV AG-879A 115/230VAC 50/60Hz 10/5A 8.7/4.3A 10 5
IS1800HGDV AG-879B 115/230VAC 50/60Hz 15/8A 15/7.8A 15 8
Note: Use only a 5 mm x 20 mm time-delay glass fuse (250V).