SVA1000X User Manual 44
frequency of the Delta Marker.
The function is invalid in Zero span.
3. M -> Start Freq
Set the start frequency of the analyzer to the frequency of the current marker.
If the Normal marker is selected, the start frequency will be set to the frequency of the
current marker.
If the Delta or Delta Pair marker is selected, the start frequency will be set to the
frequency of the Delta Marker.
The function is invalid in Zero span.
4. M -> Stop Freq
Set the stop frequency of the analyzer to the frequency of the current marker.
If the Normal marker is selected, the stop frequency will be set to the frequency of the
current marker.
If the Delta or Delta Pair marker is selected, the stop frequency will be set to the
frequency of the Delta Marker.
The function is invalid in Zero span.
5. M ->Ref Level
Set the reference level of the analyzer to the amplitude of the current marker.
If the Normal marker is selected, the reference level will be set to the amplitude of the
current marker.
If the Delta or Delta Pair marker is selected, the reference level will be set to the
amplitude of the Delta Marker.
6. ΔM->Span
Set the span of the analyzer to the frequency difference between the two markers in Delta
marker type.
If the Normal marker is selected, this function is invalid.
The function is invalid in Zero span.
7. ΔM->CF
Set the center frequency of the analyzer to the frequency difference between the two markers
in Delta marker type.
If the Normal marker is selected, this function is invalid.
The function is invalid in Zero span.
2.3.3 Marker Fn
Special marker functions including Noise Marker, N dB BW and Freq Counter.