SVA1000X User Manual 33
Speed: Activates the default fast sweep time rule.
Accuracy: Activates the normal sweep time rule to ensure increased measurement
accuracy. The Speed sweep time rule provides a fast measurement function that
decreases the sweep time. Using Fast Sweep will decrease the measurement accuracy.
2.2.4.3 Sweep
Sets the sweep mode in single or continuous, the default is continuous. The corresponding
icon of the sweep will be displayed in the status bar at the left of the screen.
1. Single
Sets the sweep mode to Single”. The number on the parameter icon denotes the current
sweep number.
2. Numbers
Sets the number of sweeps for a single sweep. In single sweep mode, the system executes
the specified number of sweeps and the number shown on the icon in the status bar at the left
of the screen varies with the process of the sweep.
3. Continue
Sets the sweep mode to Continue”. The character Cont on the parameter icon denotes the
analyzer is sweeping continuously.
If the instrument is in single sweep mode and no measurement function is enabled, press
this key and the system will enter continuous sweep mode and sweep continuously if the
trigger conditions are satisfied.
If the instrument is in single sweep mode and a measurement function is on, press this
key and the system will enter continuous sweep mode and measure continuously if the
trigger conditions are satisfied.
In continuous sweep mode, the system will send a trigger initialization signal
automatically and enter the trigger condition judgment directly after each sweep.
Table 2-17 Sweep Times
Parameter
Explanation
Default
1
Range
1 ~ 9999
Unit
N/A
Knob Step
1
Direction Key Step
1
2.2.4.4 Sweep Mode
Sweep mode includes auto, sweep and FFT operation modes.