Guidance and Manufacturer’s Declaration
This equipment is suitable for hospitals except for near active HF SURGICAL EQUIPMENT and the RF
shielded room of an ME SYSTEM.
WARNING: Use of this equipment adjacent to or stacked with other equipment should
be avoided because it could result in improper operation. If such use is necessary, this
equipment and the other equipment should be observed to verify that they are operating
normally.
WARNING: Use of accessories, transducers and cables other than those specified or
provided by the manufacturer of this equipment could result in increased electromagnetic
emissions or decreased electromagnetic immunity of this equipment and result in improper
operation.
Guidance and Manufacturer’s Declaration—Electromagnetic Emissions
This Medical-Grade UPS is intended for use in the electromagnetic environment specified below. The customer
or the user of this Medical-Grade UPS should assure that it is used in such an environment.
Standard Description Test Level/Limit Guidance
EN 55011:
2009+A1:2010
Radiated
Emissions
Models:
SMX700HG,
SMX1200XLHG
Class A Group 1,
30 - 1000 MHz
See notes 1 and 2
Model:
OMNIX350HG
Class B Group 1,
30 - 1000 MHz
See notes 1 and 3
EN 55011:
2009+A1:2010
Conducted
Emissions
Models:
SMX700HG,
SMX1200XLHG
Class A Group 1,
150 kHz – 30 MHz
See notes 1 and 2
Model:
OMNIX350HG
Class B Group 1,
150 kHz – 30 MHz
See notes 1 and 3
Notes:
1. Group 1: The Medical-Grade UPS uses RF energy only for its internal function. Therefore, its RF emissions are
very low and unlikely to cause any interference in nearby electronic equipment.
2. Class A: The Medical-Grade UPS is suitable for use in all establishments other than domestic and those
directly connected to the public low-voltage power supply network that supplies buildings used for domestic
purposes.
NOTE The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas and
hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally
required) this equipment might not offer adequate protection to radio-frequency communication services. The
user might need to take mitigation measures, such as relocating or re-orienting the equipment.
3. Class B: The Medical-Grade UPS is suitable for use in all establishments, including domestic and those
directly connected to the public low-voltage power supply network that supplies buildings used for domestic
purposes.