Taking Thickness Measurements
1. Set the Sound Velocity on the meter
2. Coat the piece to be measured with coupling agent
3. Place the probe on the area and the Coupling Indicator will appear
on the LCD
4. Read the measurement on the LCD
5. When you remove the probe the value will stay on the LCD and
the Coupling Indicator will disappear
6. Press the Save Button to save the measurement
If the Coupling Indicator ashes or doesn’t appear it means that the
coupling is not good.
Zero Calibration
1. Select the correct frequency for the probe in use
2. Set the Sound Velocity to 5900m/s
3. Select the proper Receiving Gain (see System Setup)
4. Coat the 4mm Standard Test Block with Coupling Agent and press
the probe to the Test Block
5. If the Coupling Indicator is showing on the LCD then press the
Zero Button to initiate Zero Calibration
6. The meter will beep and then the screen will indicate that the
calibration is complete
7. If Zero Calibration is not properly completed the meter will retain
the original value
8. To delete the calibration data, see Memory Manager