INSTRUMENTS
Measurement Methods
There are three base measurement methods:
• Single Measurement: Measurement at one point.
• Double Measurement: Measure one point twice. During the two
measurements, the probe’s crosstalk interlayer plate should be
placed in a perpendicular direction. Take the minimum readout as
the accurate thickness of the material.
• Multi-point Measurement: Make several measurements in a range,
and take the minimum readout as the thickness of the material.
Measurements on Cylindrical Surfaces
When measuring cylindrical material, such as pipes or oil tubes, it is
important to properly adjust the angle between the probe’s crosstalk
interlayer plate and the axial line of the material to be measured.
1. Couple the probe with the material to be measured
2. Make the probe’s crosstalk interlayer plate perpendicular or
parallel to the axial line of the object
3. Shake the probe vertically along the axial line of the object, the
readouts displayed on screen will change regularly
4. Use the minimum readout
The standard for selecting the angle between the probe’s crosstalk
interlayer plate and the axial line of the object depends on the curvature of
the object. For a pipe with a large diameter the probe’s crosstalk interlayer
plate should be perpendicular to the axial line of the object. For a pipe
with small diameter, you can measure with the probe’s crosstalk interlayer
plate being both parallel and perpendicular to the axial line of the object,
and take the minimum readout as the thickness.
REED Instruments
1-877-849-2127 | [email protected] | www.reedinstruments.com