110/113/114/115/117
Users Manual
18
MIN MAX Recording Accuracy and Response Time (113)
Specified accuracy of the measurement function ±40 counts in k CHEK for changes >500 ms in duration, ±12 counts in e for changes
>325 ms in duration. Typical 100 ms response to 80 %. Response time not specified for capacitance.
Table 8. Input Characteristics (113)
Function Input Impedance (Nominal) Common Mode Rejection Ratio
k CHEK ~3 kΩ <300 pF >60 dB at dc, 50 or 60 Hz
Open Circuit Test Voltage Full Scale Voltage
Ohms <2.7 V dc <0.7 V dc
Diode Test <2.7 V dc <2.000 V dc
Short Circuit Current
Ohms <350 μA
Diode Test <1.0 mA