Table 6. Accuracy Specifications
Function Range Resolution
Accuracy
± ([% of Reading] + [Counts])
Model
DC Millivolts 600.0 mV 0.1 mV 0.5 % + 2 110, 114, 115, 117
DC Volts
6.000 V 0.001 V
0.5 % + 2 110, 114, 115, 11760.00 V 0.01 V
600.0 V 0.1 V
DC, 45 to 500 Hz 500 Hz to 1 kHz
Auto-V LoZ
[1]
True-rms
600.0 V 0.1 V 2.0 % + 3 4.0 % + 3 114, 117
k CHEK
[4]
6.000 V 0.001 V
2.0 % + 3 4.0 % + 3 11360.00 V 0.01 V
600.0 V 0.1 V
45 to 500 Hz 500 Hz to 1 kHz
AC millivolts
[1]
True-rms
600.0 mV 0.1 mV 1.0 % + 3 2.0 % + 3 110, 114, 115, 117
AC Volts
[1]
True-rms
6.000 V 0.001 V
1.0 % + 3 2.0 % + 3 110, 114, 115, 11760.00 V 0.01 V
600.0 V 0.1 V
Continuity
[5]
600 Ω 1 Ω
Beeper on <20 Ω, off >250 Ω. Detects
opens or shorts of 500 μs or longer.
110, 114, 115, 117
--- --- 113
Ohms
[5]
600.0 Ω 0.1 Ω 0.9 % + 2
110, 113, 114, 115,
117
6.000 kΩ 0.001 kΩ 0.9 % + 1
60.00 kΩ 0.01 kΩ 0.9 % + 1
600.0 kΩ 0.1 kΩ 0.9 % + 1
110, 114, 115, 117
6.000 0Ω 0.001 MΩ 0.9 % + 1
40.00 MΩ 0.01 MΩ 5.0 % + 2
Diode Test
[5]
2.000 V 0.001 V
0.9 % + 2 115, 117
2.0 % + 3 113
Capacitance
[5]
1000 nF 1 nF 1.9 % + 2
113, 115, 117
10.00 μF0.01 μF1.9 % + 2
100.0 μF0.1 μF1.9 % + 2
9999 μF1 μF
100 μF - 1000 μF: 1.9 % +2
>1000 μF: 5 % + 20
Lo-Z Capacitance
(Power-up option)
1 nF to 500 μF 10 % + 2 typical 115, 117
AC Amps True-rms
[1]
(45 Hz to 500 Hz)
6.000 A 0.001 A
1.5 % + 3 115, 117
10.00 A
[3]
0.01 A
DC Amps
6.000 A 0.001 A
1.0 % + 3 115, 117
10.00 A
[3]
0.01 A
Loading ...