HANTEK6082BE 80MHz PC Based USB Oscilloscope

Users Manual - Page 17

For HANTEK6082BE.

PDF File Manual, 87 pages, Read Online | Download pdf file

HANTEK6082BE photo
Loading ...
Loading ...
Loading ...
Hant
ek 6052BE/6082BE
1
1
.6 Self Calibration
The self calibration routi
ne lets you optimize the oscilloscope signal path for maximum
measurement accuracy. You can run the routine at any time but you should always run
the routine if the ambient temperature changes by 5v or more. For accurate calibration,
power on the oscilloscope and wait twenty minutes to ensure it is warmed up. To
compensate the signal path, disconnect any probes or cables from the input connectors.
Then, access the “Utility -> Calibration” option and follow the directions on the screen.
The self calibration routine takes about several minutes.
.7 Function Check
Perform this
functional check to verify that your oscilloscope is operating correctly.
Connect the oscilloscope
You should connect the A-Type Plug of USB cable to your PC USB port and connect the
B-Type Plug of USB cable to oscilloscope USB port.
Input a sign
al to a channel of the oscilloscope
The oscilloscope is equipped with two channels plus external trigger.
Please input signal in the following steps:
1. Set the attenuation switch on the probe as 10X and connect the probe on the
oscilloscope with CH1. Aim the slot in the probe connector at the faucet on BNC of CH1
and insert, then, turn right to lock the probe. Finally, attach the tip of probe and ground nip
to the Connector of Probe compensator.
U
SERS MANUAL 15