Using additional functions
34 |
RS 5000
Charging the rechargeable battery using the
MCA
800 charging adapter
If you have purchased an additional stethoset receiver or an
optional MCA 800 charging adapter, you can also charge the ste-
thoset receiver using the charging adapter.
1 Select a suitable country adapter and slide it onto the power
supply unit until it clicks audibly into place.
2 Connect the connector of the power supply unit to the connec-
tion socket of the MCA 800 charging adapter.
3 Plug the power supply unit into the wall socket.
4 Connect the charging adapter to the charging contacts on the
stethoset receiver. The charging adapter is magnetically ori-
ented to the correct position.
The rechargeable battery is being charged. The LED on the
charging adapter lights up red. When the rechargeable battery is
fully charged, the LED lights up green.
If the LED on the charging adapter flashes red and green, a charging
error/battery error has occurred. In this case, disconnect the
charging adapter from the stethoset receiver. More information on
the rechargeable battery can be found on page 28.
WARNING
Interference due to magnetic fields!
The MCA 800 charging adapter generates stronger permanent mag-
netic fields that could cause interference with cardiac pacemakers
and implanted defibrillators (ICDs).
Always maintain a distance of at least 3.94” (10 cm) between the
holding magnets of the charging adapter and the cardiac pace-
maker or implanted defibrillator.
You can also use the stethoset receiver during charging.
Please note, however, that the receiver can get slightly warm
during charging. The charge status of the rechargeable bat-
tery is displayed on the transmitter (see page 16).
3 h
100 %
US
EU
UK
AU
21
3 4