Hantek MSO5000D Mixed Storage Oscilloscope

User Manual - Page 74

For MSO5000D.

PDF File Manual, 92 pages, Read Online | Download pdf file

MSO5000D photo
Loading ...
Loading ...
Loading ...
Specifications
MSO5000D Digital Storage Oscilloscope User Manual 69
Chapter 8 Specifications
8.1 Technical Specifications
All specifications herein mentioned apply to the MSO-5000D oscilloscopes. Before checking an
oscilloscope from our company to see if it complies with these specifications, make sure it meets
the following conditions:
The oscilloscope must have been operating continuously for twenty minutes under the
specified operating temperature.
The Do Self Cal operation must be performed through the Utility menu if the operating
temperature changes by more than 5.
The oscilloscope must be within the factory calibration interval.
All specifications are guaranteed unless noted ‘typical’.
Oscilloscope Specifications
Horizontal
Sample Rate Range
500MS/s dual channel
Waveform Interpolation
(sin x)/x
Record Length
Maximum 1M samples per single-channel; maximum 512K
samples per dual-channel (4K, 40K optional)
SEC/DIV Range
MSO5062D MSO5102D
MSO5202D
8ns/div to 40s/div, in a 2, 4, 8
sequence
2ns/div to 40s/div, in a 2, 4, 8
sequence
Sample Rate and
Delay Time Accuracy
±50ppm over any ≥1ms time interval
Delta Time Measurement
Accuracy
(Full Bandwidth)
Single-shot, Normal mode
± (1 sample interval +100ppm × reading + 0.6ns)
>16 averages
± (1 sample interval + 100ppm × reading + 0.4ns)
Sample interval = s/div ÷ 200
Position Range
2ns/div to 8ns/div
(-8div × s/div) to 40ms
20ns/div to 80μs/div
(-8div × s/div) to 40ms
200μs/div to 40s/div
(-8div × s/div) to 400s
Vertical
A/D Converter
8-bit resolution,
each channel sampled simultaneously
VOLTS/DIV Range
2mV/div to 10V/div at input BNC