Hantek MSO5000D Mixed Storage Oscilloscope

User Manual - Page 30

For MSO5000D.

PDF File Manual, 92 pages, Read Online | Download pdf file

MSO5000D photo
Loading ...
Loading ...
Loading ...
Basic Operation
MSO5000D Digital Storage Oscilloscope User Manual 25
Applying a window to the YT waveform changes the waveform so that the start and stop values
are close to each other, which reduces the discontinuities.
The Math FFT function has three FFT Window options. There is a trade-off between frequency
resolution and amplitude accuracy for each type of window. You shall determine which one to
choose according to the object you want to measure and the source signal characteristics.
Window
Measurement
Characteristics
Hanning
Periodic Waveform
Better frequency, poorer amplitude accuracy than Flattop
Flattop
Periodic Waveform
Better amplitude, poorer frequency accuracy than Hanning
Rectangular
Pulse or Transient
Waveform
Special-purpose window applicable to discontinuous
waveforms. This is actually the same as no windows.