There may be
condensation on the glass
door
High ambient temperature or humid
conditions
Dry with a cloth
The display shows “So”
and compressor does not
run
The temperature sensor is open circuit
Call for service
The display shows “SC”
and compressor does not
run
The temperature sensor is short
circuited.
Call for service
The display shows “HH”
and compressor does not
run
The ambient temperature is above
40°C/104°F
Stop using immediately
The display shows “LL”
and compressor does not
run
The ambient temperature is below -
40°C/-40°F.
Stop using immediately
If you continue to experience issues, contact Accucold’s technical support department at 1-888-4-
MEDLAB.