Summit ARG15PV456TEST 15 Cu.Ft. Upright Vaccine Refrigerator, Certified to NSF/ANSI 456 Vaccine Storage Standard

User Manual - Page 12

For ARG15PV456TEST.

PDF File Manual, 42 pages, Read Online | Download pdf file

ARG15PV456TEST photo
Loading ...
Loading ...
Loading ...
12
Table of Parameters and Factory Settings
Parameter
Function
Set Range
Default
Parameter
Function
Set Range
Default
Set
Setpoint
4.0°C
For glass door models
5.0°C
C3H
Product
Temp. High
Temp. Alarm
-19.9~20.0°C / -
3.8~68.0°F
8.0°°C /
46.4°F
For solid door models
PAS
Password
00~99
15
C3L
Product
Temp. Low
Temp. Alarm
-19.9~20.0°C / -
3.8~68.0°F
2.0°C /
3.6°F
E1
Lowest set
-19.9°C / 3.8°F ~ set
2.0°C / 35.6°F
C4
Hysteresis
0.1~20.0°C /
0.1°C /
point limit
temp.
Alarm
1.8~36.0°F
0.18°F
E2
Higher set
point limit
20.0°C / 68.0°F ~set
temp.
10.0°C / 50°F
C5
Start-up
Temp. Alarm
delay
00~99 min
60 min
E3
Temp.
Hysteresis
0.1~20.0°C /
0.2~36.0°F
2.0°C / 3.6°F
C6
Temp. Alarm
Delay
00~99 min
3 min
E40
Turning On
delay time
00 ~10 min.
3 min.
C7
Power Off
relay Alarm
00=do not alarm /
01= alarm
01
E41
Comp. start
delay time
00 ~10 min.
3 min.
C8
Alarm relay
close after,
muffle alarm
relay switch
00=open /
01=close
00
E5
Offset on Air
Temperature
-10.0~10.0°C / -
18.0~18.0°F
1.0°C / 1.8°F
For 1 and 3 cu. Ft.
models
2.0°C / 3.6°F
C9
Restart time
after buzzer
mute
00=do not start /
01 ~30 min=restart
time
10 min
For 6, 8, 12 and 15 cu.
Ft. models
E6
Offset on
Product
Temp.
-10.0~10.0°C / -
18.0~18.0°F
0.0°C / 32°F
C10
Comp.
Force stop
time
01~99 min
99 min
Loading ...
Loading ...
Loading ...