DO600-en-GB_V7.4 12/16
3
Operation
PreparetheElectrode
Theelectrodeisshipped“dry”andrequiresfillingwiththesuppliedelectrolytesolutionbefore
use.Themembraneshouldbeinplaceanddoesnotneedreplacement.Followtheprocedureat
theendofthisguideforfillingthebondedcap.
NOTE:Makesurethattherearenoairbubblesintheelectrodebeforeuse.
PoweringtheExStik®
TheExStik
®
usesfour(4)CR2032LithiumIonBatteries.PresstheON/OFFbuttontoturnthe
meteronoroff.Ifthebatteriesareweak,the‘BAT’indicatorappearsontheLCD.Theauto
powerofffeatureshutstheExStik
®
offautomaticallyafterapproximately10minutesof
inactivity.Theautopowerofffeaturemaybetemporarilydisabledforconvenienceorfor
extendedpolarizationtime.
StartupPolarizationPeriod
WhentheExStik
®
ispoweredupforthefirsttime,theelectroderequirespolarization.Inorder
forthistooccurathreeminute(approx.)polarizationperiodmustelapsebeforemeasurements
canbemade.
Whenthemeterisleftpoweredon,aspecialcircuitisincludedinthismeterwhichmaintainsa
verysmallbiasingcurrenttotheelectrodeforaperiodofsevendays.Thiskeepstheelectrode
polarizedandenablestheusertomakeimmediatemeasurementswithoutwaitingforthe
electrodetorepolarize.
EachtimetheExStik®isturnedon,thepolarizationtimerisresetandthe7daypolarization
periodis
initiated.Asmallasteriskatthebottomrightcornerofthedisplayisusedtoindicate
thatthepolarizationtimercircuitisactiveevenwhenthemeterisoff.
TurnOnDiagnostics
1. WhenthemeteristurnedONtheLCDdisplays“SELF”and“CAL”whilethemeterrunsa
diagnosticroutine.
2. DuringthistimethemeterisrecallingtheUserCalibrationdata,performingselfdiagnostics
&initializingthecircuitry.
3. Afterthisfunctioniscomplete,themeterproceedstothenormal
measurementmode.
4. TheDO600shouldbecalibrateddailyformeasurementaccuracy.