DO600-en-GB_V7.4 12/16
11
Appendices
TroubleshootingGuide
Symptom PossibleCauses Action
Unitwillnotpoweron‐Batteriesnotinplace
Deadbatteries
Batterypolarityincorrect
ReplaceBatteries
ReplaceBatteries
Reorient/Replace
Batteries
“BAT”indicatorshownon
display
Batteriesareweak‐ReplaceBatteries
UnstableReadings‐Insufficientelectrolytein
probe(Airbubblespresent
whenprobeisinverted.)
Electrolyteisdepleted
Replaceelectrolyte,and
membranecapassembly.
Readingsdriftdown‐Insufficientstirring(the
probeconsumesoxygenat
themeasuringsurface,
requiringconstantsample
movementacross
membrane)
Moveprobeinsampleor
stirsample
SlowResponse‐Dirtyordamaged
membrane
Replaceelectrolyteand
membranecap
Electrodecannotbe
calibrated
Depletedelectrolyte
Dirtyordamaged
membrane
Replaceelectrolyteand
membranecap
Electrodecannotbe
calibratedafterreplacing
theelectrolyteand
membranecap
Dirtyprobe(Cathodeisnot
ashinygoldcolor)
Cleancathodewith
Polishingpaper*perform
theelectronincZero
calibration
Samplereadingisfrozen‐Unitisin“HOLD”mode
Unitislocked
ReleaseHOLD
(momentarillypress
Mode/Holdbutton)
Removebatteries,press
ON/OFFbutton,replace
batteriesandrestart
*PolishingpaperisavailableintheDO603membranekit
