Summit LAB11 4.8 Cu.Ft. -85ºC Ultra Low Chest Freezer

User Manual - Page 12

For LAB11.

PDF File Manual, 16 pages, Read Online | Download pdf file

LAB11 photo
Loading ...
Loading ...
Loading ...
User Instructions Ultra Low Temperature Freezers Model LAB
Side 10 | 14
17.
D
E
F
A
U
L
T
S
E
T
T
I
N
G
V
A
L
U
E
S
Lable Name Set
Set
Set point -80
Hy
Differential
3.0
LS Minimum set point
-85
US
Maximum set point
-50
Ot
Thermostat probe calibration
5.0
OdS Outputs delay at start up 0
AC
Anti-short cycle delay
5
CCt
Continuos cycle duration 0.0
CCS
Set point for continuous cycle, 0.0
COn
Compressor ON time with faulty probe
1
COF
Compressor OFF time with faulty probe
0
CH
Kind of action CL
CF
Temperature measurement unit °C
rES
Resolution
dE
dLy
Display temperature del a y
0.0
IdF
Interval between defrost cycles
0
MdF Maximum length for defrost
0
dFd Displaying during defrost
rt
dAd
MAX display delay after defrost
0
ALc Temperat. alarms configuration
rE
ALU MAXIMUM temperature alarm
10.0
ALL
MINIMUM temperature alarm
5.0
AFH Differential for temperat. alarm recovery
2.0
ALd Temperature alarm delay
60
dAo
Delay of temperature alarm at start up
8.0
tbA Alarm relay disabling
n
AoP Alarm relay polarity (oA1=ALr)
OP
i1P Digital input polarity
CL
i1F Digital input configuration
AUS
did Digital input alarm delay
0
nPS Number of activation of pressure switch
0
odc Compress status when open door
no
rrd
Regulation restart with door open alarm
alarm
y
HES Differential for Energy Saving
0
Adr. Serial address
1
PbC Kind of probe
pt1
onF
on/off key enabling
no
rSE Room probe display
-85
rEL Software release
1.6
Ptb Map code
1
17. FACTORY SETTINGS DIXELL XR30CH