Accucold DL2KIT USB Temperature Logger Kit for Aftermarket I

User Manual - Page 8

For DL2KIT.

PDF File Manual, 12 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
R5100615
8
To Use
Once the Data Logger has been programmed, place the temperature sensor (in glycol bottle) in the location to be monitored, such as
inside a refrigerator or freezer. Data Logger may be placed on top of unit with LCD display easily visible and the alarm audible. Data Logger
displays internal temperature of unit being monitored, as well as maximum and minimum temperatures reached. The Data Logger’s
maximum and minimum readings reflect the highest and lowest temperatures since the battery was inserted or since the RESET button
was last pressed.
If the temperature measurement rises above or falls below the set temperature range, the alarm will sound. To silence the alarm, press
any key ONCE.
Data Logging / Recording Function
In the Data Logger Setting menu user confirms time, date, and sampling rate.
TO RECORD DATA