Electromagnetic Compatibility Information
Table 1
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture’s declaration – electromagnetic emissions
The SBM 03 is intended for use in the electromagnetic environment specified below.
The customer or the user of the SBM 03 should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions CISPR 11 Group 1 The SBM 03 uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions CISPR 11 Class B The SBM 03 is suitable for use in all establishments other than domestic and
those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions IEC 61000-3-2 Not applicable
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Not applicable
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer’s declaration – electromagnetic immunity
The SBM 03 is intended for use in the electromagnetic environment specified below. The customer or the user of the SBM 03 should assure
that it is used in such an environment.
IMMUNITY test IEC 60601 test level Compliance level Electromagnetic environment – guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30 %.
Power frequency (50/60 Hz)
magnetic field IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic fields should be at levels cha-
racteristic of a typical location in a typical commercial or
hospital environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.