Loading ...
Loading ...
Loading ...
21
Never hold the instrument under water as otherwise
liquid can penetrate and damage the instrument.
Never place any heavy objects on the instrument.
9. Technical details
Model no. SBM 03
Measurement
method
Oscillometric, non-invasive blood
pressure measurement on the
wrist
Measurement range Cuff pressure 0 300 mmHg,
systolic 60 – 260 mmHg,
diastolic 40 – 199 mmHg,
Pulse 40 180 beats/minute
Display accuracy systolic ± 3 mmHg,
diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Measurement inac-
curacy
Max. permissible standard devia-
tion according to clinical testing:
systolic 8 mmHg /diastolic
8 mmHg
Memory 60 memory spaces
Dimensions L 84 mm x W 62 mm x H 30 mm
Weight Approx. 112 g (without batteries)
Cuff size 14.0 to 19.5 cm
Permissible operating
conditions
+5 °C to +40 °C, ≤ 90 % relative
air humidity (non-condensing)
Permissible storage
conditions
-20 °C to +55 °C, ≤ 95 % rela-
tive air humidity, 800 1050 hPa
ambient pressure
Power supply
2 x 1,5 V
AAA batteries
Battery life For approx. 180 measurements,
depending on the blood pressure
level and/or pump pressure
Accessories Instruction for use, 2 x 1.5 V AAA
batteries, storage box
Classification Internal supply, IPX0, no AP or
APG, continuous operation, type
BF applied part
Technical information is subject to change without notifi-
cation to allow for updates.
This unit is in line with European Standard
EN 60601-1-2 and is subject to particular precautions
with regard to electromagnetic compatibility (EMC).
Please note that portable and mobile HF communica-
tion systems may interfere with this unit. More details
can be requested from the stated Customer Service
address or found at the end of the instructions for use.
Loading ...