16
5. Technical Data
Product Description: Forehead Thermometer
Model: MC-720
Sensor: Thermopile
Temperature Display: 4-digit, °F display in 0.1 degree increments
3-digit, °C display in 0.1 degree increments
Accuracy: Forehead measurement mode ±0.2°C (±0.4°F) within 35.0°C to 42.0°C (95.0°F
to 107.6°F), ±0.3°C (±0.5°F) for other range
Surface measurement mode ±0.3°C (±0.5°F) within 22.0°C to 42.2°C (71.6°F to
108.0°F). Outside this range, it is ±2°C (±3.6°F) or
±4% (percentage of actual temperature),
whichever is greater.
Room temperature mode ±2°C (±3.6°F) within 10°C to 40°C (50°F to 104°F).
Out of this range, the measurement result cannot
meet the aforementioned accuracy.
Measurement Range: Forehead measurement mode 34.0°C (93.2°F) to 42.2°C (108.0°F)
Surface measurement mode -22.0°C (-7.6°F) to 80.0°C (176.0°F)
Room temperature mode 10°C (50°F) to 40°C (104°F)
Power Supply: 3.0V DC, 1 CR2032 lithium button battery
Power Consumption: 0.018 W
Battery Life: With a new battery, approx. 2,500 measurements
Operating Temperature
and Humidity:
10°C (50°F) to 40°C (104°F), 15 to 85%RH
MC-720-AP_A_M.book Page 16 Friday, May 22, 2015 11:01 AM