4. Troubleshooting and Maintenance
14
3. There are variances in the measurement values.
Measurements were not taken at the same location.
The probe lens is dirty.
4. The temperature measured on this unit is not equivalent to oral-temperature.
Forehead thermometers may not always calculate the exact oral-equivalent value.
It is recommended to take regular measurements on subjects such as family members, etc., who are in
good physical condition, to know their normal temperature.
4.3
Maintenance
Check the unit if it is dropped. If unsure, please contact your local OMRON representative to have your unit
checked.
When the probe lens becomes dirty, wipe it lightly with a soft dry cloth or a cotton swab.
Do not wipe the probe lens with tissue paper or a paper towel.
Clean the probe lens carefully to avoid damaging it.
- Use a 70% alcohol swab or soft cotton moistened with 70% alcohol to clean the probe lens.
- Allow the probe lens to fully dry for at least 1 minute.
Do not store the unit in the following types of places. Doing so may damage the unit.
- Wet locations.
- Locations with high heat and humidity or those that are exposed to direct sunlight. Areas close to heating
equipment, dusty locations, or environments where there are high salt concentrations in the air.
- Locations where the unit will be subjected to leaning over, falling, shock, or vibration.
- Pharmaceutical storage areas or locations where corrosive gases are present.