Quick Start Guide BD / BF / ED / FD / FED (E3)
Page 6
The domestic socket must also provide a protective conductor. Make sure that the connection of the protective
conductor of the domestic installations to the chamber’s protective conductor meets the latest technology. The
protective conductors of the socket and plug must be compatible!
When connecting, please observe the regulations specified by the local electricity supply company and as well
as the VDE directives (for Germany). We recommend the use of a residual current circuit breaker.
Prior to connection and start-up, check the power supply voltage. Compare the values to the specified data
located on the chamber’s type plate.
Only use original connection cables from BINDER.
To completely separate the chamber from the power supply, you must disconnect the power plug. Install the
chamber in a way that the power plug is easily accessible and can be easily pulled in case of danger.
BF, FD, FED: Do not place the power cable over the door gap when the chamber is hot after operation.
Optional internal socket (BD, BF): Connection of appliances with protection class I. Charge max.500 W. Hous-
ing protection. to EN 60529: IP 67. Voltage 230 V at 50 Hz and at 60 Hz power frequency. Current type: 1N~
Placing the charging material in the usable volume
Do NOT place samples outside the defined usable volume. Observe the wall clearances = take at least 10% of the
interior dimensions as wall clearance.
Details on calculating the usable volume can be found in the operating manual in chapter 16.2.
Do NOT load the usable volume by more than half to enable sufficient airflow inside the chamber
Do NOT divide the usable volume into separate parts with large area samples.
Do NOT place samples too close to each other in order to permit circulation between them and thus obtain a
homogenous distribution of temperature.