Loading ...
Loading ...
Loading ...
Quick Start Guide BD / BF / ED / FD / FED (E3)
Page 5
Requirements for the location of installation
Installation inside a building on anon-flammable surface
Do NOT set up chambers in non-ventilated recesses. Ensure sufficient ventilation for dispersal of the heat.
Permissible ambient temperature range during operation: +18 °C up to +40 °C / 64.4 °F up to 104 °F.
Permissible ambient humidity: 70 % r.H. max., non-condensing
Installation height: max. 2000 m / 6562 ft. above sea level.
Do not install or operate the chamber in potentially explosive areas.
Power supply fluctuations must not exceed ± 10 % of the nominal voltage
Electrical connection data and installation
The chambers and come with an IEC connector plug with a shockproof plug. Only use original connection ca-
bles from BINDER.
Pollution degree (acc. to IEC 61010-1): 2. Over-voltage category (acc. to IEC 61010-1): II.
Housing protection (acc. to 60529): IP 20
Chamber size
56
115
260
720
BD
Nominal voltage (+/-10%) at 50 and 60 Hz
V
230
230
230
230
Current type
1N~
1N~
1N~
1N~
Nominal power
kW
0.30
0.35
0.85
1.65
Nominal power with option internal socket
kW
0.80
0.85
1.35
2.15
Chamber fuse 5x20 mm / 250V / time-lag (T)
A
6.3 external
6.3 external
8.0 external
12.5 external
Overtemperature protective device class 1
°C
120
120
120
120
BF
Nominal voltage (+/-10%) at 50 and 60 Hz
V
230
230
230
230
Current type
1N~
1N~
1N~
1N~
Nominal power
kW
0.40
0.40
0.90
1.75
Nominal power with option internal socket
kW
0.90
0.90
1.40
2.25
Chamber fuse 5x20 mm / 250V / time-lag (T)
A
6.3 external
6.3 external
8.0 external
12.5 external
Overtemperature protective device class 1
°C
120
120
120
120
ED
Nominal voltage (+/-10%) at 50 and 60 Hz
V
230
230
230
400
Current type
1N~
1N~
1N~
3N~
Nominal power
kW
1.05
1.25
2.25
4.10
Chamber fuse 5x20 mm / 250V / time-lag (T)
A
6.3 external
6.3 external
12.5 external
---
Miniature circuit breaker
A
---
---
---
16 internal
Overtemperature protective device class 1
°C
330
330
330
330
FD, FED
Nominal voltage (+/-10%) at 50 and 60 Hz
V
230
230
230
400
Current type
1N~
1N~
1N~
3N~
Nominal power
kW
1.10
1.30
2.30
4.50
Chamber fuse 5x20 mm / 250V / time-lag (T)
A
6.3 external
6.3 external
12.5 external
---
Miniature circuit breaker
A
---
---
---
16 internal
Overtemperature protective device class 1
°C
330
350
330
330
Loading ...