Declaration – electromagnetic immunity
The Philips Avent breast pump is intended for use in the electromagnetic
environment specied below. The user of the Philips Avent breast pump
should assure that it is used in such an environment.
Immunity
test
IEC
60601
test
level
Com-
pli ance
level
Electromagnetic environment – guidance
Conducted
RF IEC
61000-4-6
3 Vrms
150 kHz
to 80
MHz
3 Vrms Portable and mobile RF communications equipment should be
used no closer to any part of the Philips Avent breast pump,
including cables, than the recommended separation distance
calculated from the equation applicable to the frequency of the
transmitter.
Recommended separation distance d=1,17 P The breast pump
can be turned o.
Radiated RF
IEC 61000-
4-3
3 V/m
80 MHz
to 2,5
Ghz
3 V/m 80 MHz to 800 Mhz d=1,17 P
800 MHz to 2,5 GHz d=2,33 P
where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer and d is
the recommended separation distance in metres (m).
Field strengths from xed RF transmitters, as determined
by an electromagnetic site survey, should be less than the
compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked
with the following symbol:
Caution: Although the breast pump is compliant with applicable
EMC directives, it may still be susceptible to excessive
emissions and/or may interfere with other equipment. A
consequence can be that the breast pump turns o or the
on/o button starts ashing (see troubleshooting for more
information), which is acceptable as this will not lead to
unacceptable risks. To prevent interference, keep other electric
equipment away from the breast pump during expressing.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic
propagation is aected by absorption and reection from structures, objects
and people.
Field strengths from xed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to xed RF
transmitters, an electromagnetic site survey should be considered. If the
measured eld strength in the location in which the Philips Avent breast
pump is used exceeds the applicable RF compliance level above, the Philips
Avent breast pump should be observed to verify normal operation.
39