Technical information
EMC information
The Philips Avent breast pump needs special precautions regarding EMC
and needs to be installed and put into service according to the EMC
information provided in this section.
Portable and mobile RF communications equipment can aect the Philips
Avent breast pump.
The electronic breast pump has no essential performance. Due to EMC
inuences, like wireless home network devices, mobile phones or cordless
telephones, the breast pump can turn o or go into error mode. This will
not lead to unacceptable risks. To prevent interference, keep other electric
equipment at least 3.3 feet/1.0 m away from the breast pump during
expressing and do not stack with other electrical equipment. Cable length of
the adapter: 2.50 meter/ 8.2 feet.
Electromagnetic Compatibility (EMC, IEC 60601-1-2)
Declaration – electromagnetic emissions
The Philips Avent breast pump is intended for use in the electromagnetic
environment specied below. The user of the Philips Avent breast pump
should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions CISPR
11
Group 1 The Philips Avent breast pump uses RF energy only for its
internal function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby electronic
equipment.
RF emissions CISPR
11
Class B
The Philips Avent breast pump is suitable for use in all
establishments, including domestic establishments and those
directly connected to the public low voltage power supply
network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage
uctuations/ icker
emissions IEC
61000-3-3
Complies
37
English