Loading ...
Loading ...
Loading ...
PT3 device is intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled. The customer or
the user of the PT3 device can help prevent electromagnetic
interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters)
and the PT3 as recommended below, according to the maximum
output power of the communication sequipment.
Rated maximum
output power of
transmitter W
150kHz to
80MHz
80MHz to
800MHz
800MHz to
2,5GHz
Recommended separation distances between portable and
mobile RF communications equipment and the PT3 device.
0.01
0.1
1
10
100
0.12
0.38
1.2
7.8
12
0.12
0.38
1.2
7.8
12
0.23
0.73
2.3
3.3
23
d=1.2 p d=1.2 p d=2.3 p
Separation distance according to frequency
of transmitter m
Electrostatic
discharge(ESD)
lEC61000-4-2
WARRANTY
Please contact your dealer or the device center in case of a
claim under the warranty. If you have to send in the unit,
enclose a copy of your receipt with clear statement of defect
description.
The warranty terms as below:
1. The warranty period for device is one year from date of
delivery. In case of a warranty claim, the date of delivery
has to be proven by means of the sales receipt or invoice.
2. Repairs under warranty do not extend the warranty period.
3. The following cases are excluded under the warranty
4. Liability for direct or indirect consequential losses
caused by the unit is excluded even if the damage to the
unit is accepted as a warranty claim.
All damage which has arisen due to improper
treatment, e.g. nonobservance of the user instruction.
All damage which is due to repairs or tampering by the
customer or unauthorized third parities.
Damage which has arisen during transport from the
manufacturer to the consumer or during transport to the
service centre.
Accessories which are subject to normal wear and tear.
Manufactured for iHealth Labs, Inc.
USA:
iHealth Labs, Inc.
www.ihealthlabs.com
120 San Lucar Ct., Sunnyvale, CA 94086, USA
+1-855-816-7705
Famidoc Technology Co., Ltd.
No. 212 Yilong Road, Hexi Industrial Zone,
Jinxia, Changan Town, Dongguan 523853,
Guangdong Province, China
Europe:
iHealthLabs Europe SAS
www.ihealthlabs.eu
36 rue de Ponthieu, 75008, Paris, France
iHealthLabs Europe SAS
www.ihealthlabs.eu
36 rue de Ponthieu, 75008, Paris, France
0120
Immunity test Compliance
level
ConductedRF
lEC6 1000-4-6
Radiated RF lEC
61000-4-3
3V rms
150 kHzto
80 MHz
3V rms
3V/m
3V/m
80 MHz to
2.5 GHz
±6kV
contact
±8k Vair
Portable and mobile RF
communications
equipment should be
used no closer to any
part of the PT3 device,
than the recommended
separation distance
calculated from the
equation applicable to
the frequency of the transmitter.
Recommended separation distance.
where P is the maximum output power rating of the transmitter in
watts (W) according to the. transmitter manufacturer and d is the
recommended separation distance in meters(m). Field strengths
from xed RF transmitters, as determined by an electromagnetic
site survey, a. should be less than the compliance level in each
frequency range.
b. Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE I At 80 MHz end 800 MHz. the higher frequency range
applies.
NOTE2 These guidelines may not apply in all situations.
Electromagnetic propagation is aected by absorption and
reection from structures, objects and people.
a. Field strengths from xed transmitters, such as base stations
for radio(cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to xed RF transmitters, an
electromagnetic site survey should be considered. If the
measured eld strength in the location in which the PT3
device is used exceeds the applicable RF compliance level
above, the should be observed to verify normal operation. If
abnormal performance is observed, additional measures may
be necessary, such as reorienting or relocating the PT3.
b. Over the frequency range 150 kHz to 80 MHz, eld strengths
should be less than [Vi] V/m.
IEC 60501
Test level
Electromagnetic
environment - guidance
d=1.2 p
d=1.2 p
80MHz800MHz
d=1.2 p
800MHz2.5GHz
For transmitters rated at a maximum output power not listed
above, recommended separation distance d in meters(m) can
be estimated using the equation applicable to the frequency of
the transmitter, where P is the maximum output power rating of
the transmitter in watts(W) accordable to the transmitter
manufacturer.
NOTEI At 80 MHz and 800 MHz. the separation distance for the
higher frequency range applies.
NOTE2 These guidelines may not apply in all situations.
Electromagnetic propagation is aected by absorption and
reection from structures, objects and people.
10 11
Loading ...
Loading ...
Loading ...