Binder BF 115 Standard-Incubators with forced convection

User Manual - Page 78

For BF 115.

PDF File Manual, 107 pages, Read Online | Download pdf file

BF 115 photo
Loading ...
Loading ...
Loading ...
BD / BF / ED / FD / FED (E3.1) 03/2019 page 78/106
Chamber size
FED 56
FED 115
FED 260
FED 720
Different electrical data for FED-UL constructed for the USA and Canada
(model versions FED056UL-120V, FED115UL-120V, FED260UL-240V, FED720UL-208V)
Nominal voltage
(+/-10%)
at 50 Hz power
frequency
V 120 120 240 208
at 60 Hz power
frequency
V 120 120 240 208
Current type
1N~
1N~
2~
3N~
Nominal power
kW
1.20
1.40
2.50
4.50
Power plug of the power cable
NEMA
5-15P
5-15P
6-20P
L21-20P
Chamber fuse (external)
5x20 mm / 250V / time-lag (T)
A 12.5 12.5 --- ---
Miniature circuit breaker (combina-
tion element, external)
A --- --- 16 ---
Miniature circuit breaker (internal)
A
---
---
---
16
Environment-specific data
Noise level (mean value)
dB (A)
43
43
43
43
Energy consumption at 150 °C /
302 °F
Wh/h
300 340 420 800
*** Data not yet determined
All technical data is specified for unloaded chambers with standard equipment at an ambient temperature
of +22 °C +/- 3°C / 71.6 °F +/- 5.4 °F and a power supply voltage fluctuation of +/-10%. Technical data is
determined in accordance to BINDER Factory Standard Part 2:2015 and DIN 12880:2007.
All indications are average values, typical for chambers produced in series. We reserve the right
to change technical specifications at any time.
If the chamber is fully loaded, the specified heating up times may vary according to the load.
16.9 Equipment and options (extract)
To operate the chamber, use only original BINDER accessories or accessories / components
from third-party suppliers authorized by BINDER. The user is responsible for any risk arising
from using unauthorized accessories.
BD
BF
ED
FD
FED
Standard equipment
Microprocessor temperature controller
One timer function: Delayed Off
Three timer functions: Delayed On, Delayed Off and
Temperature dependent Delayed Off
Adjustable ramp function
Temperature safety controller class 3.1 acc. to DIN
12880:2007
Temperature safety controller class 2 acc. to DIN 12880:2007
Inner glass door
USB interface to read out the measured values
Communication interface Ethernet
Exhaust duct, internal diameter 50 mm / 1.97 inches, with
adjustable ventilation slide
Adjustable air change by means of rear exhaust duct (50
mm)
2 racks, chrome-plated
Loading ...
Loading ...
Loading ...