Binder MKT 115 dynamic climate chamber with expanded low-temperature range

User Manual - Page 70

For MKT 115.

PDF File Manual, 163 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
MK / MKT (E5) 08/2020 page 70/162
11. Notification and alarm functions
11.1 Notification and alarm messages overview
11.1.1 Notifications
Notifications are indicated by information icons displayed in the screen header in Normal display
An information icon serves as an indication of a certain condition.
If this condition persists, in some cases an alarm will be triggered after a fix or configurable interval. As long
as the condition persists, the information icon therefore continues to be displayed also in state of alarm. If
during alarm the conditions ends, e.g., if during a tolerance range alarm the actual value returns to within
the tolerance range, the information icon disappears, whereas the alarm will continue until manual acknowl-
edgement.
Press the flash icon next to the information icon to access the corresponding text information.
Normal display showing the text information.
The currently valid information texts are highlighted
in black (example: “Idle mode”)
Condition
Information
icon
Text information
Start after
condition occurred
1-hour preheating phase, no cooling function
“Preheating
phase”
immediately
Open chamber door
“Door open” immediately
The controller is in Idle mode (chap. 5.4).
“Idle mode immediately
Operation line “Anti-condensation” on. Anti-
condensation protection is activated.
“Anti-condensa-
tion”
immediately
Operation line “Compr. air dryer” on. Op-
tional compressed air dryer activated.
Turn operation line off when not using the
compressed air dryer!
“Compressed Air
Dryer
immediately
Notifications are not shown in the event list.