Binder B 28

User Manual - Page 14

For B 28.

PDF File Manual, 39 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
B / E 07/2017 page 14/39
Permissible ambient temperature range during operation: +18 °C up to +40 °C / 64.4 °F up to 104 °F.
At elevated ambient temperature values, fluctuations in temperature can occur.
The ambient temperature should not be substantially higher than the indicated ambient
temperature of +22 °C +/- 3 °C / 71.6 °F ± 5.4 °F to which the specified technical data
relate. Deviations from the indicated data are possible for other ambient conditions.
Permissible ambient humidity: 70 % r.H. max., non-condensing.
Installation height: max. 2000 m / 6562 ft. above sea level.
When placing several chambers of the same size side by side, maintain a minimum distance of 250 mm /
9.84 in between each chamber. Wall distances: rear 100 mm / 3.94 in, sides 160 mm / 6.29 in. Spacing
above the chamber of at least 100 mm / 3.94 in must also be maintained.
The devices are not suitable for stacking.
To completely separate the chamber from the power supply, you must disconnect the power plug. Install
the chamber in a way that the power plug is easily accessible and can be easily pulled in case of danger.
For the user there is no risk of temporary overvoltages in the sense of EN 61010-1:2010.
Do not install or operate the chamber in hazardous locations.
DANGER
Explosion hazard.
Danger of death.
Do NOT operate the chamber in potentially explosive areas.
KEEP explosive dust or air-solvent mixtures AWAY from the vicinity of the chamber.