Loading ...
Loading ...
Loading ...
Quick Start Guide BD / BF / ED / FD / FED (E3)
Page 10
Cleaning and decontamination
DANGER
Electrical hazard! Danger of death.
Ø Do NOT spill water or cleaning agents over the inner and outer surfaces.
Disconnect the chamber before cleaning. Disconnect the power plug.
Completely dry the chamber before turning it on again.
CAUTION
Contact with skin, ingestion. Skin and eye damage due to chemical burns.
The recommended neutral cleaning agent may cause health problems in contact with skin and if
ingested.
In case of eye contact the recommended disinfectant spray may cause eye damage due to chemi-
cal burns.
Ø Do not ingest. Keep away from food and beverages.
Wear protective gloves (butyl or nitrile rubber, penetration time >480 minutes) and sealed
protective goggles
Avoid skin contact with the neutral cleaning agent.
For cleaning wipe the surfaces with a moistened towel or use cleaning agents and disinfectants.
For surface protection, perform cleaning as quickly as possible.
After cleaning completely remove cleaning agents from the surfaces with a moistened towel. Allow the
chamber to dry.
Following use of disinfectants: allow the chamber to dry, and aerate it sufficiently.
Exterior surfaces,
racks,
door gaskets
Standard commercial cleaning detergents free from acid or halides.
Alcohol-based solutions.
Recommended for cleaning: neutral cleaning agent, art. no. 1002-0016.
Instrument panel Standard commercial cleaning detergents free from acid or halides.
Recommended for cleaning: neutral cleaning agent, art. no. 1002-0016.
Zinc coated hinge parts
rear chamber wall
Standard commercial cleaning detergents free from acid or halides.
Do NOT use a neutral cleaning agent on zinc coated surfaces.
Inner chamber Standard commercial cleaning detergents free from acid or halides.
Standard commercial surface disinfectants free from acid or halides.
Alcohol-based solutions.
Recommended for cleaning: neutral cleaning agent, art. no. 1002-0016.
Recommended for chemical disinfection: disinfectant spray, art. no. 1002-0022.
Decontamination by hot air sterilization: The drying and heating ovens ED, FD and FED can be hot air steri-
lized at 190 °C / 374 °F for at least 30 minutes. All inflammable goods must be removed from the interior.
With the incubators BD and BF it is possible to perform a hot-air disinfection at 100 °C / 212 °F.
All inflammable goods must be removed from the interior.
Decontamination by autoclaving: You can remove and sterilize the racks.
Loading ...
Loading ...
Loading ...