16 ellume·lab User Manual
10 DiagnosticTesting
10.1 Icons
10.2 DiagnosticTestPreparation
1. Ensure the ellume·lab is turned on and charged to a minimum of 20% battery
2. Tap the Patient Test or Control Test icon from the home screen and follow the
on-screen instructions. More information can be found in the printed
Instructions for Use (IFU) and Quick Reference Instructions (QRI) in each test
carton.
* The remaining testing process has been clinically validated for each ellume·lab
test and corresponding external controls.
Caution:
• Do not press the power button while testing.
• Only remove the eStick and test kit components from the packaging
immediately before testing.
• The eStick must be used within 60 minutes from insertion into the ellume·lab.
The eStick will expire and become unusable if no sample is loaded within 60
minutes from insertion.
• Keeptheellume·labatwhiletestingisinprogress.Rotatingthedevicewhile
performing a test will prompt a warning message.
• Do not expose the ellume·lab to unstable lighting conditions while testing. If a
warningisdisplayed,removetheellume·labandeStickfromuctuatinglight
sources.
Patient Test Control Test
Previous Results Training
Flat surface
Position while
performing test