English Instruction Sheet
Page 4
Sourcing mA
Use source mode whenever you need to supply current into a
passive circuit. Insert the test leads into the terminals as shown
below.
Note
A path must exist for current to flow between the output
+ and mA terminals, otherwise the display will indicate
an overload (OL) when you set an output value.
OFF
COM
+
705
LOOP CALIBRATOR
SOURCE
SIMULATE
MEASURE
VDC
mA
SpanCheck
Loop Power
Source
Press to toggle
Source or Simulate.
Press to step output
up/down in 25 %
steps.
Press to scroll
up/down 0.01 mA.
Hold down to scroll
faster.
For 0.001 mA steps,
see Pushbutton
Functions.
Press simultaneously
once for slow ramp
twice for fast ramp
three times for
stepped ramp.
or
rt04f.eps