FLUKE BT520 Battery Analyzer

User Manual for the Fluke BT520 Battery Analyzer - Page 37

For BT520.

PDF File Manual, 62 pages, Read Online | Download pdf file

BT520 photo
Loading ...
Loading ...
Loading ...
Use the Product
Make Measurements 3
3-11
Set Measurement Thresholds
The Product lets you define upper and lower measurement thresholds or tolerance
ranges. These defined threshold values are then compared to the measured values to
automatically identify and prompt the user with a PASS, FAIL or WARN indicator of
battery out of tolerance conditions.
The threshold function is disabled by default. You can set up to 10 set of thresholds and
select one threshold as needed.
To set and select measurement thresholds:
1. On the measurement screen, push the More softkey and Threshold softkey to open
the Select Threshold menu.
2. Use and to select one threshold set out of ten.
hsn072.jpg
3. Use and L to highlight the value to be edited among Voltage lower, Reference,
Warning, Fail, Low limit, and Notes.
4. Edit the selected field.
a. Use the and + softkeys to change the values for Warning and Fail.
b. For other fields, push the Edit softkey, use the arrow keys to edit the value, and
then push the Confirm softkey to save the value.
5. Once all threshold values are correct, push the Confirm softkey on the save the
threshold set.
The threshold set is applied and the T-X (X stands for value of No.) icon and the
corresponding PASS/WARN/FAIL indication shows on the display.
To disable measurement thresholds:
1. On the measurement screen, push the Threshold softkey to open the Select
Threshold menu.
The value of No. is already highlighted.
2. Use to set No. to ---.
3. Push the Confirm softkey.
The T-X icon no longer shows on the display.