FLUKE 87V/IMSK Industrial Multimeter Service Combo Kit with current clamp, 1000 V, 10 A

User Manual for the Fluke 87V/IMSK Industrial Multimeter Service Combo Kit with current clamp, 1000 V, 10 A - Page 20

For 87V/IMSK.

PDF File Manual, 58 pages, Read Online | Download pdf file

87V/IMSK photo
Loading ...
Loading ...
Loading ...
80 Series V
Users Manual
14
When measuring voltage, the Meter acts approximately
like a 10 M (10,000,000 ) impedance in parallel with
the circuit. This loading effect can cause measurement
errors in high-impedance circuits. In most cases, the error
is negligible (0.1% or less) if the circuit impedance is
10 k (10,000 ) or less.
For better accuracy when measuring the dc offset of an ac
voltage, measure the ac voltage first. Note the ac voltage
range, then manually select a dc voltage range equal to or
higher than the ac range. This procedure improves the
accuracy of the dc measurement by ensuring that the
input protection circuits are not activated.
MIN MAX
RANGE
REL
Hz %
AutoHOLD
Peak MIN MAX
4½ DIGITS
1 Second
˚
C/
˚
F
OFF
mA
A
mV
V
V
A
A
mA
COM
V
400mA
FUSED
10A MAX
FUSED
A
TRUE RMS MULTIMETER
87
V
LOLO
MIN MAX
RANGE
REL
Hz %
AutoHOLD
Peak MIN MAX
4½ DIGITS
1 Second
˚
C/
˚
F
OFF
mA
A
mV
V
V
A
A
mA
COM
V
400mA
FUSED
10A MAX
FUSED
A
TRUE RMS MULTIMETER
87
V
LOLO
Switch Box
V
+
AC Voltage
DC Voltage
V
mV
aom2f.eps
Figure 2. Measuring AC and DC Voltage