Changing the Auto Set Options
6
93
Changing the Auto Set Options
With the next procedure you can choose how auto set
behaves when you press the AUTO-MANUAL (auto set) key.
1
USER
Display the USER key labels.
2
F1
Open the USER OPTIONS menu.
3
ENTER
Open the AUTO SET ADJUST menu.
If the frequency range is set to > 15 Hz, the Connect-and-
View function responds more quickly. The response is
quicker because the test tool is instructed not to analyze
low frequency signal components. However, when you
measure frequencies lower than 15 Hz, the test tool must
be instructed to analyze low frequency components for
automatic triggering:
4
ENTER
Select 1 Hz and up, then jump to
Input Coupling:
When you press the AUTO-MANUAL (auto set) key, the input
coupling can either be set to dc or left unchanged:
5
ENTER
Select Unchanged.
When you press the AUTO-MANUAL (auto set) key glitch
capture can either be set to On or left unchanged:
6
ENTER
Select Unchanged.
Note
The auto set option for the signal frequency is
similar to the automatic trigger option for the
signal frequency. (See Chapter 4: “Automatic
Trigger Options”). However, the auto set option
determines the behavior of the auto set function
and shows only effect when you press the auto
set key.