ENGLISH 9
The software’s measurement features will work right out of
the box. However, if you want the most precise measurements
possible, use a calibration slide (not included) to more precisely
calibrate the software to the microscope’s view.
1. Place a calibration slide on the microscope stage. Make sure
the slide is level and in focus.
2. Set the camera resolution to the highest setting
(2592x1944).
3. Click the Calibration icon to open the calibration tool. A
line/measurement tool will appear on the live stream and a
window will open.
4. In the window, adjust the Magnification setting to reflect the
objective lens you are using.
5. Adjust the Ruler Unit setting to reflect the units of measure
on your calibration slide.
6. Using the mouse, grab the edges of the line/measurement
tool. Drag the line to lengthen or shorten it as needed so that
the line spans across a portion of the calibration ruler.
7. Enter the length of the line you created in the box labeled
Actual Length.
8. Your microscope is calibrated. You can now replace the
calibration slide with a specimen slide. The microscope will
remain calibrated as long as you do not change the objective
lens.
7. CALIBRATING THE IMAGER