FLUKE TI480 60HZ Thermal Imaging Camera with SuperResolution, 640x480

User Manual for the Fluke TI480 60HZ Thermal Imaging Camera with SuperResolution, 640x480 - Page 30

For TI480 60HZ.

PDF File Manual, 55 pages, Read Online | Download pdf file

TI480 60HZ photo
Loading ...
Loading ...
Loading ...
Ti200, Ti300, Ti400, Ti450, Ti480
Users Manual
26
Auto Capture
Start Capture
Use Auto Capture settings to capture and save an infrared
image, or series of images, automatically.
Interval
Sets the number of hours, minutes, or seconds between image
captures.
Note
The minimum interval available can be affected by
the file type and visible light camera settings. Some
combinations create larger file sizes that take longer
to capture and save and create a higher minimum
interval compared to others.
Image Count
Sets a number of images to capture. Or, select Maximum
Memory to capture and save images until the chosen storage
memory is full or the battery runs out of power.
Manual
Trigger
Select to automatically capture images when Start Capture is
selected.
Temp Trigger
Select to capture images when a value is above or below a set
temperature limit when Start Capture is selected.
Set Temp
Trigger
With Temp Trigger selected, set the temperature and conditions
to trigger the auto capture of images.
Wireless
Bluetooth
Uses Bluetooth technology to connect the Imager to a device
such as a wireless headset. See
Wireless Connectivity.
WiFi Hotspot
Uses the Imager to create a wireless Hotspot when no WiFi
network exists. See
Wireless Connectivity.
WiFi Network
Connects the Imager to a WiFi network so you can sign into your
Fluke Connect account on the Imager. See
Wireless
Connectivity.
Table 12. Camera Menu (cont.)