Addendum
Adjust the Sensor and prism so that the laser beam strikes the prism and is
reected back into the Sensor.
Warning
Do not stare into the beam.
1. Keep the prism dust cap on. If the Sensor and prism are roughly positioned
with each other during mounting, the laser beam should strike the prism dust
cap and is readily visible. If the beam is so far o󰀨 target that the beam misses
the prism completely, hold a sheet of paper in front of the prism to locate the
beam and readjust it.
2. Make these adjustments:
Vertical adjustment: slide the prism or Sensor up and down along the
support posts. Use the thumbwheel [
A] on the side of the prism housing.
To move the Sensor, loosen the yellow knobs [
B].
The lever [
C] must always be in the horizontal position except for mounting
and dismounting.
Figure 7-9. Horizontal and vertical adjustment of the laser beam
1
2
3