True-rms Digital Multimeters
Detailed Specifications
75
Frequency Counter Sensitivity
Approximate Voltage Sensitivity
(rms sine wave)
[1]
Input Range
15 Hz to 100 kHz
AC Bandwidth
[2]
Approximate DC
Trigger Levels
DC Bandwidth
[2]
50 mV 5 mV 1 MHz 5 mV & 20 mV 600 kHz
500 mV 25 mV 1 MHz 20 mV & 60 mV 1 MHz
5 V 0.25 V 700 kHz 1.4 V & 2.0 V 80 kHz
50 V 2.5 V 1 MHz 0.5 V & 6.5 V 1 MHz
500 V 25 V 300 kHz 5 V & 40 V 300 kHz
1000 V 50 V 300 kHz 5 V & 100 V 300 kHz
Approximate Current Sensitivity
(rms sine wave)
Input Range
15 Hz to 10 kHz
AC Bandwidth
Approximate DC Trigger
Levels
DC Bandwidth
500 μA 25 μA
100 kHz
5000 μA 250 μA
100 kHz
50 mA 2.5 mA 100 kHz
400 mA 25 mA 100 kHz
5 A 0.25 A 100 kHz
10 A 1.0 A 100 kHz
NA NA
[1] Maximum input = 10 x range (1000 V maximum, 2 x 10
7
V-Hz product maximum). Noise at low frequencies and amplitudes may affect accuracy.
[2] Typical frequency bandwidth with full scale (or maximum 2 x 10
7
V-Hz product) rms sine wave.