FLUKE 289-KIT5 289 True RMS Industrial Data Logging Multimeter Kit - Includes FREE Products with Purchase

User Manual for the Fluke 289 True RMS Industrial Data Logging Multimeter Kit - Includes FREE Products with Purchase - Page 76

For 289-KIT5.

PDF File Manual, 87 pages, Read Online | Download pdf file

289-KIT5 photo
Loading ...
Loading ...
Loading ...
True-rms Digital Multimeters
Detailed Specifications
67
Electromagnetic Compatibility Standards (EMC)
European EMC....................................................EN61326-1
Australian EMC ...................................................; N10140
US FCC ...............................................................FCC CFR47: Part 15 CLASS A
Certifications .........................................................UL, CE, CSA, ; (N10140),
š
Detailed Specifications
Accuracy:
Accuracy is specified for a period of one year after calibration, at 18 °C to 28 °C (64 °F to 82 °F), with relative humidity to 90 %. Accuracy
specifications are given as: ±( [ % of reading ] + [ number of least significant digits ] ). Accuracy specification assumes ambient temperature
stable at ±1 °C. For ambient temperature changes of ±5 °C, rated accuracy applies after 2 hours. To obtain full accuracy in DC mV,
Temperature, Ohms and Lo (50) Ohms, let the meter stabilize 20 minutes after using LoZ.
True-rms:
AC mV, AC V, AC μA, AC mA, and AC A specifications are ac-coupled, true rms, and are specified from 2 % of range to 100 % of range,
except 10 A range is specified from 10 % to 100 % of range.
Crest Factor:
Accuracy is specified with AC crest factor 3.0 at full-scale, increasing linearly to 5.0 at half-scale, except the 1000 V range, where it is 1.5
at full scale, increasing linearly to 3.0 at half-scale and 500 mV and 5000 μA, where it is 3.0 at 80 % of full scale, increasing linearly to 5.0
at half-scale. For non-sinusoidal waveforms add ±(0.3 % of range and 0.1 % of reading).
AC Floor:
When the input leads are shorted together in the ac functions, the Meter may display a residual reading up to 200 counts. A 200 count
residual reading will cause only a 20 count change for readings at 2 % of range. Using REL to offset this reading may produce a much larger
constant error in later measurements.
AC+DC:
AC+DC is defined as
22
dcac +
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com
Loading ...
Loading ...