393/393 FC
Users Manual
20
AC Voltage
Range.............................................................. 600.0 V
1000 V
Resolution........................................................ 0.1 V (600.0 V)
1V (1000 V)
Accuracy.......................................................... 1 % RD + 5 digits (20 Hz to 500 Hz)
DC Voltage
Range.............................................................. 600.0 V
1500 V
Resolution........................................................ 0.1 V (600.0 V)
1V (1500 V)
Accuracy.......................................................... 1 % RD + 5 digits
mV dc
Range.............................................................. 500.0 mV
Resolution........................................................ 0.1 mV
Accuracy.......................................................... 1 % RD + 5 digits
Amps Frequency: Jaw
Range.............................................................. 5.0 Hz to 500.0 Hz
Resolution........................................................ 0.1 Hz
Accuracy.......................................................... 0.5 % RD + 5 digits
Trigger Level.................................................... 5 Hz to 10 Hz, 10 A
10 Hz to 100 Hz, 5A
100 Hz to 500 Hz, 10 A
Amps Frequency: Flexible Current Probe
Range.............................................................. 5.0 Hz to 500.0 Hz
Resolution........................................................ 0.1 Hz
Accuracy.......................................................... 0.5 % RD + 5 digits
Trigger Level.................................................... 5 Hz to 20 Hz, 25 A
20 Hz to 100 Hz, 20 A
100 Hz to 500 Hz, 25 A
Voltage Frequency
Range.............................................................. 5.0 Hz to 500.0 Hz
Resolution........................................................ 0.1 Hz
Accuracy.......................................................... 0.5 % RD + 5 digits
Trigger Level.................................................... 5 Hz to 20 Hz, 5 V
20 Hz to 100 Hz, 5 V
100 Hz to 500 Hz, 10 V