FLUKE TI95 9HZ General Use Thermal Imager

User Manual for the Fluke Ti95 9hz General Use Thermal Imager - Page 36

For TI95 9HZ.

PDF File Manual, 68 pages, Read Online | Download pdf file

TI95 9HZ photo
Loading ...
Loading ...
Loading ...
Ti90, Ti95, Ti100, Ti105, Ti110, Ti125, TiR105, TiR110, TiR125
Users Manual
30
Transmission/Transmittance Adjustment
When you do infrared inspections through infrared-transparent windows
(IR windows), not all of the infrared energy emitted from the objects of interest
is transmitted through the optical material in the window. If the transmission
percentage of the window is known, you can adjust this percentage in the
Imager or in the SmartView
software. Adjustment of the transmission
correction can make the accuracy of the temperature measurement better in
many situations.
To adjust the transmission percentage:
1. Push .
2. Push / to highlight Measurement.
3. Push or to view the menu.
4. Push / to highlight Transmission.
5. Push or to view the menu.
6. Push / to adjust the percentage.
7. Push:
to set the change and go back to the live view.
or to set the change and go back to the previous menu.
to cancel the change and go back to the live view.
Note
If Display Information is set to Display All, you see the information
about current transmission correction as
τ = xx.
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com
Loading ...
Loading ...
Loading ...