Reflected Background Temperature Compensation
Compensation for reflected background temperature is set in the Background
tab. Very hot objects or very cold objects can affect the apparent temperature
and measurement accuracy of the target or object of interest, especially when
surface emissivity is low. Adjustment of the reflected background temperature
can make the temperature measurement better in many situations. For more
information, see Emissivity Adjustment.
To adjust the background temperature:
1. Push .
2. Push / to highlight Measurement.
3. Push or to view the menu.
4. Push / to highlight Background.
5. Push or to view the menu.
6. Push / to change the value.
7. Push:
to set the change and go back to the live view.
or to set the change and go back to the previous menu.
to cancel the change and go back to the live view.
Note
If the Display is set to Display All, you see the information about
current reflected background temperature as BG = xx.x.