FLUKE TI300-60HZ Ti300 60Hz Thermal Imaging Camera, 43200 PIxels 240 x 180

User Manual for the Fluke Ti300 60Hz Thermal Imaging Camera, 43200 PIxels 240 x 180 - Page 16

For TI300-60HZ.

PDF File Manual, 54 pages, Read Online | Download pdf file

TI300-60HZ photo
Loading ...
Loading ...
Loading ...
Ti300 PRO, Ti300+, Ti400 PRO, Ti401 PRO, Ti450 PRO, Ti450 SF6, Ti480 PRO
Users Manual
12
Measurement Menu
Table 8 lists the options in the Measurement menu.
Table 8. Measurement Menu
Option Menu Option Description
Range <options>
Select the temperature range from one of the preset
measurement ranges or to a fully automatic range.
Set Level/Span
Auto
Sets the Level/Span to adjust automatically or manually.
Manual
Set Level/Span
With Level/Span set to Manual, changes the Level/Span. See
Level/Span.
Line Temp <options> Turns on/off the Line Temp.
Emissivity
Adjust Number
Sets a custom emissivity value when a value from the standard
emissivity table is not appropriate for the measurement. See
Emissivity Adjustment.
Select Table
Select an emissivity value from a list of common materials. See
Emissivity Adjustment.
Background <options>
Changes the background temperature to compensate for
reflected background temperature.
Very hot objects or very cold objects can affect the apparent
temperature and measurement accuracy of the target,
especially when surface emissivity is low. Adjust the reflected
background temperature to improve the accuracy of the
measurement.
Note
If Display is set to Display All, the background temperature
shows as BG: xx.x on the display.
Transmission <options>
Changes the values used to calculate the temperature based
on the transmission percentage of the infrared-transparent
window (IR window), relative humidity, and the distance the
Imager is from the target. See Transmission.
Note
If Display is set to Display All, the transmission correction
shows as T: xxx% on the display.