NORMA 4000/5000
Operators Manual
5-10
Measurement with Shunt
The connecting leads to the shunts should be as short as possible in order to prevent noise
voltages (see Figure 5-6).
XW Warning
To avoid possible electric shock or personal injury:
• Do not touch sense terminals. The sense terminals at the
shunts are powered with the same voltage as the power
connections.
• Shunts are not isolated. Never touch the sense terminals at
the shunts.
• Risk of injury when touching connections, internal circuits
and measuring devices that are not connected to earth
ground.
• Always adhere to the instructions regarding the sequence of
connection (see Chapter 5, “Connecting Sequence”).
HI
VOLTAGE
1000 V max
LO
P R OB E
EX T.SHUN T
10 V max
123
HI
CURRENT
10 A max
LO
ALL INPUTS MAX 1000V C ATII T O
P R OB E
EX T.SHUN T
10 V max
HI
CURRENT
10 A max
LO
HI
VOLTAGE
1000 V max
LO
L1
L2
L3
LOHI
LOHI
Max. 10 m
Load
Guard
*1
Guard
*1
V
23
V
13
esn013.eps
Figure 5-6. Aron Circuit-Shunt Measurement