FLUKE 718 30G 718-30US Pressure Calibrator with Pressure Vacuum Pump, -12 to 30 PSI

User Manual for the Fluke 718-30US Pressure Calibrator with Pressure Vacuum Pump, -12 to 30 PSI - Page 8

For 718 30G.

PDF File Manual, 31 pages, Read Online | Download pdf file

718 30G photo
Loading ...
Loading ...
Loading ...
Pressure Calibrator
Safety Information
3
Table 2. Safety Information
XW Warning
To avoid possible electric shock or personal injury:
Never apply more than 30 V between the mA terminals, or between either of the mA terminals and earth
ground.
Do not use the Calibrator to make measurements in a CAT II, CAT III, or a CAT IV environment.
CAT I equipment is designed to protect against transient from high-voltage, low-energy sources, such
as electronic circuits or a copy machine
Remove the test leads from the Calibrator before you open the battery door.
Make sure the battery door is closed and latched before you operate the Calibrator.
Do not operate the Calibrator if it is damaged.
Do not operate the Calibrator around explosive gas, vapor, or dust.
When using probes, keep fingers behind the finger guards on the probes.
Use only two 9 V batteries, properly installed in the Calibrator case, to power the Calibrator.
Follow all equipment safety procedures.
Turn off circuit power before connecting the Calibrator mA and COM terminals in the circuit. Place
Calibrator in series with the circuit.
When servicing the Calibrator, use only specified replacement parts.
Do not allow water inside the case.